![SRS Labs QMS 100 Series Скачать руководство пользователя страница 190](http://html1.mh-extra.com/html/srs-labs/qms-100-series/qms-100-series_user-manual_1354225190.webp)
3-66 Error Byte Definitions
SRS QMS Gas Analyzer
Error Byte Definitions
The
Error Bytes
described in this section store the results of the firmware-driven checks built into the
RGA ECU. Use the Error Reporting commands to query the value of the bytes.
Important
: No errors are present as long as all bits in the Error Bytes are cleared.
The
RGA Windows
software supports all the Error Reporting commands and reports the errors
detected based on their Error Codes.
The RGA Troubleshooting chapter identifies the different problems based on their Error Codes and
suggests possible causes and solutions.
STATUS Error Byte
: General Status byte. Each bit corresponds to a different type of
internal check.
Bit
Internal Check
(a)
ERROR Byte
(c)
Query
(d)
7
not used
N.A.
N.A.
6
24V External P/S
(b)
PS_ERR
EP?
5
Electrometer
(b)
DET_ERR
ED?
4
Quadrupole Mass Filter
(b)
QMF_ERR
EQ?
3
Electron Multiplier
CEM_ERR
EM?
2
not used
N.A.
N.A.
1
Filament
FIL_ERR
EF?
0
Communications
RS232_ERR
EC?
(a) RGA component checked. Each check involves several tests on the component.
(b) Checks automatically performed upon a power-on-reset
(c) Error bytes that store the results of the tests for each type of internal check.
(d) Error Reporting command that queries the error byte.
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......