![SRS Labs QMS 100 Series Скачать руководство пользователя страница 251](http://html1.mh-extra.com/html/srs-labs/qms-100-series/qms-100-series_user-manual_1354225251.webp)
B-6 Description of Schematics
SRS QMS Gas Analyzer
Description of Schematics
Schematic name: QMSE_T1
Microprocessor
An MC68HC11E9 microcontroller is used to control the system and to communicate with the host
computer. This central processing unit (CPU) also has RAM, ROM, EEPROM, UART, octal 8-bit
A/D converter, counter timers, and a multiplexed address/data bus to accommodate an external 32Kx8
RAM. The ROM is used for program storage, the RAM for data storage, the EEPROM contains
calibration values for the particular unit.
The power-up reset and power-fail interrupts are provided to the CPU by U100, a MAX705.
Digital I/O Ports
The microprocessor’s I/O ports are assigned as follows:
PA0
-RTS
Low indicates that the host computer is ready for RS232 data.
PA1
-BUSY
Low indicates that the 16-bit signal A/D converter is busy.
PA2
CAL_EN
High enables modification of EEPROM calibration data.
PA3
LE_DAC18
Set high to latch SPI data into 18-bit (RF control) D/A converter.
PA4
-CS_DAC8
Strobe low to assert CS to octal 8-bit DAC.
PA5
LD_LEDS
Strobe high to load LED data into output register.
PA6
LD_MPX
Strobe high to load multiplexer data into output register.
PA7
LD_MISC
Strobe high to load miscellaneous data into output register.
PB
0-7
A8-15
High address bits for external RAM
PC
0-7
A0-7 & D0-7
Multiplexed address and data bits for external RAM
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......