3-68 Error Byte Definitions
SRS QMS Gas Analyzer
QMF_ERR Error Byte:
Quadrupole Mass Filter RF P/S Error Byte.
Bit
Description
Error Code
7
RF_CT exceeds (V_EXT- 2V) at M_MAX
RF7
6
Primary current exceeds 2.0A
RF6
5
Not used
4
Power supply in current limited mode.
RF4
3
Not Used
2
Not Used
1
Not Used
0
Not Used
CEM_ERR Error Byte:
Electron Multiplier Error Byte.
Bit
Description
Error Code
7
No Electron Multiplier Option installed
EM7.
6
Not Used
5
Not Used
4
Not Used
3
Not Used
2
Not Used
1
Not Used
0
Not Used