![SRS Labs QMS 100 Series Скачать руководство пользователя страница 157](http://html1.mh-extra.com/html/srs-labs/qms-100-series/qms-100-series_user-manual_1354225157.webp)
RGA Command Set 3-33
Technical Reference
Command excecution times vary depending on the pre-existing ionizer conditions. The end of the
command excecution is prompted to the host computer sending out the STATUS byte over RS232.
Parameters:
VFparam, param: 0-150
: The parameter represents the magnitude of the focus plate bias potential in
Volts (
The actual bias voltage is negative
). The STATUS byte is transmitted at the end of command
execution.
VF*
: The default Focus plate biasing Voltage value is used to excecute the command. Default: 90 (V).
VF?:
Query the focus plate biasing voltage setting.
Error checking:
Number parameters must be integers within the accepted range.
The absence of a parameter (i. e. VF) is treated as a bad-parameter error.