SRS QMS Gas Analyzer
Safety
Line Voltage
The QMS system is specified for power line of either 110 V / 60 Hz or 220 V / 50 Hz. The
diaphragm pump will only operate on the specified voltage. Operating at other voltages will
damage the motor. For 110 V operation use one 3 A fuse. For 220 V operation, two 1.5 A fuses
must be used in the power entry module.
Exhaust
As shipped, the QMS system exhausts to the atmosphere. If the system is analyzing hazardous
gases, the user must make provisions to handle the exhaust from the system. A standard 1/4 inch
connection is provided for this purpose.
Ventilation
The QMS system
requires
forced air cooling to operate at a reasonable temperature. Do not
block the air inlet or exhaust on the back of the unit. Components will fail without this cooling.
Elastomer Seals
Silicone has been reported to react adversely and irreversibly with the glass contained in an
electron multiplier. In systems containing an RGA w/electron multiplier, do not use silicone
greases or oils on seals; use only hydrocarbon based materials.
Lifting
The QMS system is heavy. Two people are recommended for lifting the system. The handle is
provided for occasional use and short periods. Do not hoist or hang the system by the handle.
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......