General Utilities 2-19
SRS QMS Gas Analyzer
3.
Use the Next Item or Previous Item from the View menu to view the
sequential logs (The time and date of the scan appears on each log).
Browsing Through the Gas Library
Library Browsing Description
There are several ways to browse through the gas library depending on the display mode and
whether the Library Search utility is active. The Library display mode and search utility are
linked together to provide an intuitive interface to locate and view any gas in the library file.
Library Display mode only:
1.
Activate the Library mode by selecting the Library command from the Mode
menu.
2.
Use the Next Item and Previous Item in the View menu to plot the next and
previous gases in the list (you may also use the left and right arrow keys).
Library Search utility only:
You may start this utility from any display mode
1.
Start the search utility by selecting the Library Search command from the
Utilities menu.
2.
Press the Show All button to list all available gases.
3.
Use the scroll bar, and/or the keyboard to search the list or use the Search
Masses section to look for specific gases.
Library Display mode AND Search Utility:
The combination of these two modes provides a very quick way to look for gases.
1.
Make sure both the Library mode and the Library Search utility are active (see
procedure above).
2.
If you want to plot a gas, double click on its name in the Search utility dialog
box or click on it once (to highlight it) and press the plot button. The gas
fragment pattern will then appear as a histogram in the Library display.
Analyzing the Mass Spectrum
Spectrum Analysis description
This utility provides the user with immediate analysis of residual gas based on a complete
histogram or analog scan. Using a matrix inversion technique, the composition of the residual
gas is analyzed and the best approximation to its composition is given in either pressure units or
percentages. The user can enable up to 12 common gases for analysis.
Note: Analysis mode is only available in Analog mode or Histogram mode. The mass range must
be at least from 1 to 50 amu.
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......