3-10 Programming the RGA ECU
SRS QMS Gas Analyzer
-Check the quality of the serial connection to the host computer.
-Check the user’s communication software to make sure it is communicating properly with
the RGA.
-Check the serial numbers of the RGA ECU’s connected to the computer's serial ports.
Programming the Ionizer
Positive ions are produced in the ionizer by bombarding residual gas molecules with electrons derived
from a heated filament. The operational parameters that affect the efficiency of the ionizer are: electron
energy, ion energy, electron emission current, and focusing voltage. The
Ionizer Control
Commands
program all the ionizer voltages, turn the filament on/off, and Degas the ionizer. The STATUS byte is
transmitted at the end of their execution.
Important
: Remember that the repeller grid and the focus plate are only biased while the filament is
emitting electrons.
Setting the ionizer parameters
The following list of commands will set the electron energy to 70 eV, ion energy to high (12eV), focus
voltage to -100V and it will then turn the filament on with an emission current setting of 1.0 mA. It is a
typical example of the set of commands sent to the RGA while programming the ionizer.
EE70 Electron Energy = 70 eV. Echoes the STATUS Byte.
IE1
Ion Energy = high (12eV). Echoes the STATUS Byte.
VF100 Focus Voltage= -100V. Echoes the STATUS Byte.
FL1.0 Filament emission = 1 mA and turn filament on. Echoes the STATUS Byte.
Ionizer Programming Tips
•
It is possible to trigger scans and single mass measurements if the filament is not turned on.
•
Make sure the pressure is under 10
-4
Torr before turning the filament on or degassing.
•
Whenever possible replace a Degas procedure with a thorough Probe Bakeout (See RGA
Maintenance Chapter for details). Keep the degas time to the minimum value that will get
the job done. Frequent degassing compromises the lifetime of the filament.
•
The filament is frequently tested by the CPU while it is turned on. Consult the Filament,
Leak and Burnt LED’s after activation for immediate status feedback.
•
Wait for the STATUS byte to be returned between commands and check it every time for
possible errors.
•
In cases were a hardware error is reported retry the command one more time before
declaring a hardware problem.
•
Default parameter settings are available for all the Ionizer Control commands
•
Use the FL0 command to turn off the filament.
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......