3-50 RGA Command Set
SRS QMS Gas Analyzer
different from that of the Bayard Alpert gauge readings.
Expect to see deviations between the two
gauges as the composition of a residual gas changes.
Parameters
:
TP0:
TP_Flag is cleared. Total Pressure measurement is disallowed and a null current value is returned
as a response to a total Pressure Measurement request (note that this includes the total pressure
measurement requests at the end of scans!).
TP1:
TP_Flag is set. Total pressure measurement is fully enabled.
TP?:
Total Pressure query. A total pressure measurement is triggered and a total ion current value is
returned over RS232. The actual response to the command depends on the status of TP_Flag as described
above.
Error Checking:
The absence of a parameter is considered an error.