![SRS Labs QMS 100 Series Скачать руководство пользователя страница 191](http://html1.mh-extra.com/html/srs-labs/qms-100-series/qms-100-series_user-manual_1354225191.webp)
Error Byte Definitions 3-67
Technical Reference
PS_ERR Error Byte
: 24V P/S Error Byte.
Bit
Description
Error Code
7
External 24V P/S error: Voltage >26V.
PS7
6
External 24V P/S error: Voltage <22V.
PS6
5
Not used
4
Not Used
3
Not Used
2
Not Used
1
Not Used
0
Not Used
DET_ERR Error Byte
: Electrometer Error Byte.
Bit
Description
Error Code
7
ADC16 Test failure.
DET7
6
DETECT fails to read +5nA input current
DET6
5
DETECT fails to read -5 nA input current
DET5
4
COMPENSATE fails to read +5nA input
current
DET4
3
COMPENSATE fails to read -5nA input
current
DET3
2
Not Used
1
OP-AMP Input Offset Voltage out of range
DET1
0
Not Used