3-30 RGA Command Set
SRS QMS Gas Analyzer
Command excecution times vary depending on the pre-existing ionizer conditions. The end of the
command excecution is prompted to the host computer by sending out the STATUS byte over RS232.
Note:
The Electron Impact Ionization Energy is set to the default value when the unit is turned on.
Important:
The repeller grid and the focus plate are only biased while the filament is emitting electrons.
Parameters:
EEparam, param: 25-105:
The parameter represents the electron impact ionization energy in eV.
EE*
: The default Electron Energy value is used to run the command.
Default: 70 (eV)
EE?:
Query. The electron energy parameter setting is returned over RS232 in ASCII format.
Error checking:
Number parameters must be within the accepted range and must be integers.
The absence of a parameter (i. e. EE) is treated as a parameter-error.
FLparam, param: 0.00 - 3.50, *, ?
Description
: Electron emission current (mA).
Echo
: STATUS error byte or query response.
Set the electron emission current level in the ionizer. The parameter represents the desired electron
emission current in units of mA.
When a finite emission current is requested, the RGA biases the ionizer’s repeller grid and focus plate, and
activates the filament’s heater until the requested electron current is established.
A null parameter value, turns off the filament and grounds the ionizer’s repeller grid and focus plate.
Command excecution times vary depending on the pre-existing ionizer conditions. The end of the
command excecution is prompted to the host computer sending out the STATUS byte over RS232.
The Filament LED reflects the ionizer’s emission status at all times.
A firmware-driven “
Filament Protection Mode”
monitors the performance of the filament while it is
emitting electrons and if a problem is detected at any time, the heater is immediately shut down and the
problem is reported through the error bytes and the error LED’s (see below).
WARNING: The pressure in the chamber must be under 10-4 Torr before turning the filament
on!
Important:
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......