3-64 RGA Command Set
SRS QMS Gas Analyzer
Error checking:
The absence of a parameter (i. e. RI) is treated as an error in the parameter.
This parameter is protected by an internal calibration jumper (JP100) and a Protection-violation error will
result if the jumper is in the Calibration Disabled mode (see CE command).
RSparam, param: 600.0000 - 1600.0000, *, ?, none
Description
: RF_Driver output @ 128 amu (Peak Position Tuning command).
JP100 Jumper protected.
Echo:
Query Response.
Warning:
Please read the Peak Tuning Section of the RGA Tuning Chapter before using this command.
Program the output of the RF_Driver @ 128 amu during a Peak Posion Tuning Procedure. The parameter
(one of four peak tuning parameters) represents the voltage output selected for the RF_Driver @ 128
amu, in mV.
The magnitude of the RF determines the mass-to-charge ratio of the ions that can pass through a
quadrupole mass filter without striking the rods (i.e with stable oscillations). A linear relationship between
mass and RF amplitude is one of the most attractive features of these type of filters.
The regulated output of the RF source that powers the RGA’s quadrupole rods is controlled by, and
linearly related to, the voltage output of an RF_Driver circuit. Clearly, a linear relation exists between the
output of the RF_Driver, the RF amplitude on the rods, and the mass setting of the filter. The purpose of a
Peak Position Tuning Procedure is to determine the voltages that the RF Driver must output at 0 and 128
amu so that all the peaks in an analog spectrum appear in the right position. The RS command is used to
program the voltage output of the RF_Driver @ 128 amu during the Peak Position Tuning Procedure. The
value is saved in the non-volatile memory of the RGA ECU and used by the firmware to generate the
internal scan parameters used to step the RF during scans and single mass measurements .
Note to Supervisors
: A calibration disable jumper (JP100) available on the circuit board can be used by
a supervisor to block any attempt to modify the value of the DI parameter. Supervisors may use this
feature to prevent accidental changes in the calibration parameters by inexperienced operators. Setting
JP100 will disable peak tuning of the ECU.
Parameters
:
RSparam, param: 600.0000 - 1600.0000:
If Calibration is enabled by the JP100 jumper (See CE
Command), the parameter is saved into the non-volatile memory of the RGA ECU and the internal scan
parameters used to step the RF during scans and single mass measurements are updated accordingly.
RS*
: Use this format to replace the peak tuning parameter with the original factory setting for the
RF_Driver output @ 128 amu. The factory value is retrieved from memory, and used as the new
parameter value to excecute the command as above.
RS?
: Query. Returns over RS232 the parameter value currently saved in memory.
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......