Windows Software Error Codes
4-9
Technical Reference
Troubleshooting:
Contact SRS.
DET5
Type of Error:
Electrometer
Error Message:
Electrometer Error: DETECT fails to read -5nA input current.
Error Cause:
The logarithmic output of the picoammeter is not within the levels expected for a
-5 nA input current.
Troubleshooting:
Contact SRS.
DET6
Type of Error:
Electrometer
Error Message:
Electrometer Error: DETECT fails to read +5nA input current.
Error Cause:
The logarithmic output of the picoammeter is not within the levels expected for a
+5 nA input current.
Troubleshooting:
Contact SRS.
DET7
Type of Error:
Electrometer
Error Message:
Electrometer Error: ADC16 Test Failure.
Error Cause:
The 16 bit analog-to-digital converter used to digitize the logarithmic output of
the picoammeter reports an output greater than 15 mV when its input is
grounded.
Troubleshooting:
Contact SRS.