3-22 Programming the RGA ECU
SRS QMS Gas Analyzer
3.
The Burnt and Leak LED’s indicate specific filament problems and are turned on, in
addition to the Error LED, whenever the ionizer’s emission is internally shut down or not
established as requested.
4.
Error Queries
: Queering the
Error Bytes
with the
Error Reporting
commands.
The “Error Byte Definitions” section of this chapter describes the different error bytes used
to store the results of the internal checks. The “Error Reporting Commands” query the error
bytes and are used to quickly diagnose problems.
Important
: The
RGA Windows
software supports all the Error Reporting commands and routinely
monitors the Error Bytes. Detected errors are immediately reported to the user and identified based on
Error Codes. See the RGA Troubleshooting chapter for recommended solutions to the problems.
The
STATUS error byte
is the gateway to error diagnosis:
Important: The RGA is free of detected errors as long as the STATUS byte is
clear (no bits set).
Each bit of the STATUS byte reflects the result of a different type of
internal check
. Each internal
check involves
several
different
tests
on a component of the RGA (See the RGA Troubleshooting
Chapter for details). The results of the specific tests are stored in check-specific
error bytes
. All error
bytes can be queried by the computer, using the Error Reporting commands, to find out which bits are
set.
STATUS Bit
Internal Check
(a)
ERROR Byte
(c)
Query
(d)
7
not used
N.A.
N.A.
6
24V External P/S
(b)
PS_ERR
EP?
5
Electrometer
(b)
DET_ERR
ED?
4
Quadrupole Mass Filter
(b)
QMF_ERR
EQ?
3
Electron Multiplier
CEM_ERR
EM?
2
not used
N.A.
N.A.
1
Filament
FIL_ERR
EF?
0
Communications
RS232_ERR
EC?
(a) RGA component checked. Each check involves several tests on the component.
(b) Internal checks automatically performed upon a power-on-reset
(c) Error bytes that store the results of the tests for each type of internal check.
(d) The Error Reporting command that queries the error byte.
Содержание QMS 100 Series
Страница 5: ...v SRS QMS Gas Analyzer...
Страница 79: ...Technical Reference Manual QMS 100 Series Gas Analyzer Version 3 2 1 2012...
Страница 102: ......
Страница 124: ......
Страница 194: ......
Страница 216: ...5 8 Accessing Internal Components SRS QMS Gas Analyzer Internal View Map of Internal Components...
Страница 238: ......