RGA Command Set 3-65
Technical Reference
RS:
Uses the current parameter value to recalculate the internal scan parameters used to step the RF
during scans and single mass measurements. This is often used to compensate against small temperature
drifts in the mass scale, caused by drifts in the output of the RF_Driver.
Error checking:
The absence of a parameter (i. e. RS) is treated as an error in the parameter.
This parameter is protected by an internal calibration jumper (JP100) and a Protection-violation error will
result if the jumper is in the Calibration Disabled mode (see CE command).