Command lists
10.3 Command groups of the SIWAREX WP251
SIWAREX WP251
184
Manual, 12/2015, A5E37203357A
1127
Manual emptying
OFF
The command can be issued only in weighing step 0. The "Emptying
signal" bit is then set to FALSE and a linked digital output, if any, is reset.
The command is used to clean the scale or to close it manually after
service.
1128
Rest emptying
The command can only be accepted if an emptying option (emptying
based on Limit 3 or time) is configured in DR 25.
Within a weighing cycle, the command causes the coarse and fine signal
to be shut off immediately and the scale jumps directly to the emptying
step. A log is not created and a tolerance check is not performed.
If the command is issued outside the dosing cycle (in weighing step 0),
the emptying signal is set for the specified time – if time-based emptying
was configured in DR 25. If "Emptying based on Limit 3" is configured,
the emptying signal is switched until Limit 3 is fallen below.
1141
Continue weigh-
ing
If a dosing cycle is in the "stopped" state (for example, before creation of
a log in NAWI mode), the dosing cycle can be continued using this com-
mand.
10.3
Command groups of the SIWAREX WP251
The following commands can be triggered in the scale data block DB_SCALE in the area
CMD1 to CMD3:
Table 10- 7 Command groups of the SIWAREX WP251
Command code Description
1 ... 1999
The meanings of the commands correspond to the command
2000 + X
Reading of a data record, X corresponds to the data record number.
Example: Transfer data record 3 from the SIWAREX module to the SIMATIC CPU
→ 2000
+ 3 = command code 2003
4000 + X
Writing of a data record, X corresponds to the data record number.
Example: Transfer data record 3 from the SIMATIC CPU to the SIWAREX module
→ 4000
+ 3 = command code 4003
7001
Read all data - Read all data from the SIWAREX to the CPU
7002
Write all data - Write all data from the CPU to the SIWAREX (service mode has to
be turned on)
Additional information on transmission of commands from the control program by means of
the SIMATIC interface is available in chapter Integration in SIMATIC S7-1200 (Page 185).
Содержание SIWAREX WP251
Страница 28: ...Mounting 5 3 Mounting on the SIMATIC S7 1200 SIWAREX WP251 28 Manual 12 2015 A5E37203357A ...
Страница 40: ...Connection 6 12 Attachment of calibration protection plate SIWAREX WP251 40 Manual 12 2015 A5E37203357A ...
Страница 56: ...Commissioning 7 5 Service with the SIWATOOL program SIWAREX WP251 56 Manual 12 2015 A5E37203357A ...
Страница 162: ...Scale parameters and functions 8 38 DR 48 date and time 2 for Modbus SIWAREX WP251 162 Manual 12 2015 A5E37203357A ...
Страница 212: ...Technical specifications 13 3 Approvals SIWAREX WP251 212 Manual 12 2015 A5E37203357A ...
Страница 218: ...List of abbreviations B 1 List of abbreviations SIWAREX WP251 218 Manual 12 2015 A5E37203357A ...