Command lists
10.2 Command lists
SIWAREX WP251
180
Manual, 12/2015, A5E37203357A
A data or command error is signaled if a command cannot be executed or if the sent data
record is rejected.
Table 10- 1 Service commands
Com-
mand
code
Command
Description
Pro-
tected
1
Service mode ON
Turn on service mode
2
Service mode OFF
Turn off service mode
3
Weight Simulation on
Turn on test mode. The simulation value from data record 16 is used in-
stead of the measured value for calculation of the process values. Simula-
tion mode must be enabled generally in DR 3 beforehand.
4
Weight Simulation off
Switch off test mode.
11
Load factory settings
The command resets the SIWAREX to the "ex works" status. During this
process:
•
All parameters and saved data (including log memory, logbook, IP
addresses and Modbus addresses)
•
All message buffers (diagnostic buffer, trace memory, etc.) are reset
•
A configured recovery point, if any, is overwritten with default settings.
P
12
Load standard parame-
ters
Like "Load factory settings" (command code 11), but interface settings for
Ethernet and Modbus RTU are not reset to the factory setting.
P
31
Load recovery parameter The last recovery point created with command 51 is loaded.
P
51
Create recovery parame-
ter
Backs up all scale parameters as a recovery point that can then be loaded
when needed using command 31.
P
60
Set Calibration Point 0
Calibration point 0 valid / save values for calibration point 0. The current
load cell signal or the currently measured digits are assigned to calibration
weight 0 from DR 3 and entered as "Calibration digit 0" in DR 3.
P
61
Set Calibration Point 1
Calibration point 1 valid / save values for calibration point 1. The current
load cell signal or the currently measured digits are assigned to calibration
weight 1 from DR 3 and entered as "Calibration digit 1" in DR 3.
P
62
Set Calibration Point 2
Calibration point 2 valid / save values for calibration point 2. The current
load cell signal or the currently measured digits are assigned to calibration
weight 2 from DR 3 and entered as "Calibration digit 2" in DR 3.
P
81
Shift Characteristics
Move calibration characteristic. The command defines the current weight
of the scale as the new zero point (0 kg) and shifts the entire characteristic
curve without changing the slope. The command can be used, for exam-
ple, in order to compensate parts used for mounting calibration weights on
the scale at the end of the calibration.
P
82
Automatic calibration
Calculation of scale characteristic using the load cell parameters from data
record 10. The calculated characteristic is entered directly in data records
3 and 4 and is thus immediately active following execution of the com-
mand. The scale must be empty when the command is executed. The
accuracy of a scale that is automatically calibrated depends heavily on the
mechanical setup!
P
Содержание SIWAREX WP251
Страница 28: ...Mounting 5 3 Mounting on the SIMATIC S7 1200 SIWAREX WP251 28 Manual 12 2015 A5E37203357A ...
Страница 40: ...Connection 6 12 Attachment of calibration protection plate SIWAREX WP251 40 Manual 12 2015 A5E37203357A ...
Страница 56: ...Commissioning 7 5 Service with the SIWATOOL program SIWAREX WP251 56 Manual 12 2015 A5E37203357A ...
Страница 162: ...Scale parameters and functions 8 38 DR 48 date and time 2 for Modbus SIWAREX WP251 162 Manual 12 2015 A5E37203357A ...
Страница 212: ...Technical specifications 13 3 Approvals SIWAREX WP251 212 Manual 12 2015 A5E37203357A ...
Страница 218: ...List of abbreviations B 1 List of abbreviations SIWAREX WP251 218 Manual 12 2015 A5E37203357A ...