When working with the LISST-Holo2, you will often find the instrument in sleep mode.
The instrument enters sleep mode when waiting to sample, when a sampling program
has finished, or after 10 minutes of inactivity.
If the endcap LED does not flash green every 5 second, the instrument is in sleep
mode.
To wake the LISST-Holo2 from sleep, move the Switch Lever to the
‘1’ position,
then back to the
‘0’ position.
Moving to the
‘1’ position will wake the instrument. Moving quickly back to the ‘0’
position will keep the instrument from starting the current sampling program.
The LISST-Holo2 will boot up. The endcap LED will flash yellow for approximately 40 to
60 seconds. When it is ready, the LED will blink green every 5 seconds. See section 10
for more information about the LED colors and status light indicators.
NOTE: When external power is provided, the LISST-Holo2 will run off external power
instead of the internal batteries. If external power is disconnected, the LISST-Holo2 will
automatically switch to battery power (assuming the batteries have sufficient charge).
3.3 Take a hologram
If you have followed the instructions in section 3.2 the LISST-Holo2 is now powered up
and connected to your PC. It is time to take a hologram!
Open up a web-browser on your computer, e.g. Firefox, Google Chrome or Internet
Explorer/Edge.
To connect to the LISST-Holo2, enter URL
address bar.
Note: The IP address of the instrument can be changed, and in most cases it must be
changed when connecting to a local network
– see your local system administrator for
more information on how to get a static IP address. See the FAQ in section 9 for more
information on how to change the IP address of the LISST-Holo2.
The LISST-Holo2 home page will then show.
To take a hologram, click the
Tools
option.