8
LISST-Holo2
User’s Guide
3
Getting Started
3.1 Introduction
This section is designed to provide the first-time user simple instructions to quickly
learn the operation of the LISST-Holo2. It will walk the user through setting up and
capturing hologram within a few minutes after opening the case for the first time. The
experienced user may find this section useful for a quick review of the various
procedures.
Section 4 describes how to install and use the software for processing the raw
holograms. It is not necessary to install any software to go through this section and
become familiar with the instrument and its operation.
Section 5 discusses how to configure and deploy the instrument in the field including
charging the internal batteries.
3.2 Connect cables and power
up
To program and offload data from the LISST-Holo2 you need a PC with a web browser
(e.g. Firefox, Chrome, Edge) and access to a power source. The LISST-Holo2 contains
internal batteries, however, for this tutorial we’ll demonstrate how to use the instrument
with external power.
Take out the instrument stands and place the LISST-Holo2 on the stands.
Take out the clear plastic box and open it up.
Inside you will find a few cables and power supplies, described below.
The blue cable is an Ethernet cable used for communication with the LISST-Holo2.
The Ethernet cable comes with an attached coupler that allows you to extend the cable
by plugging into a second Ethernet cable. The coupler is provided for convenience and
it’s not required for use.
The black cable is a 5-pin cable used to power the LISST-Holo2 when working in the
laboratory.
The 5-pin cable carries
power
to the LISST-Holo2 from the included 110/220VAC
power supply; it is
NOT
used for communication or battery charging. The USB
connector on this cable is for accessing the embedded Linux system, which should only
be done under instruction from Sequoia Scientific, Inc.
The blue Ethernet cable must be used for communication.
Look at the LISST-Holo2 connector endcap.
Remove the connector caps and connect the 5-pin cable to the 5-pin bulkhead
connector (just below the serial number) and the blue Ethernet cable to 8-pin bulkhead
connector (just above the serial number).
Note the white plastic vent cover next to the 5-pin connector, which is used during
battery charging.
NEVER OPEN THE VENT EXCEPT AS DESCRIBED IN SECTION
5.2. LEAVING IT OPEN RISKS FLOODING AND DESTROYING THE
INSTRUMENT.
Содержание LISST-Holo2