Pattern Trigger
The Pattern trigger identifies a trigger condition by looking for a specified pattern. The
pattern trigger can be expanded to incorporate delays similar to other triggers. Pattern
durations are evaluated using a timer. The timer starts on the last edge that makes the
pattern “true”. Potential triggers occur on the first edge that makes the pattern false,
provided that the time qualifier criterion has been met.
The oscilloscope
provides 4
patterns: logical AND, OR, NAND and NOR combination of the channels. Each channel
can set to low, high or invalid.
Channel 1High
Channel 2 Low
AND Waveform
Trigger position 1
Trigger position 2
Do the following steps to set pattern trigger:
1. Press the
Setup
button on the front panel to enter the TRIGGER function menu.
2. Press the
Type
softkey; turn the
Universal Knob
to select Pattern, and then push
down the knob to confirm.
3. Press each channel softkey to select
Invalid
,
High
or
Low
.
Low
sets the pattern to low on the selected channel. A low is a voltage level that is
less than the channel's trigger level or threshold level.
High
sets the pattern to high on the selected channel. A high is a voltage level that is
greater than the channel's trigger level or threshold level.
Invalid
sets the pattern to don't care on the selected channel. Any channel set to
don't care is ignored and is not used as part of the pattern.
However, if all channels in the pattern are set to
Invalid
, the oscilloscope will not
trigger.
Adjust the trigger level for the selected analog channel by turning the Trigger
Level knob.
Invalid
doesn’t need to set trigger level.
4. Press the
Next Page
softkey to enter the second page of the pattern trigger menu.
5. Press the
Logic
softkey and then turn the
Universal Knob
to select the desired
logic combination
AND
,
OR
,
NAND
or
NOR
.
6. Press the
Time
softkey; then turn the
Universal Knob
to select the desired
time value.
7. Press the
Holdoff Close
softkey to turn on the Holdoff function; then turn the
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...