SPI Triggering
This part will provide a brief introduction and description for the operation of the SPI trigger.
1. Press
Setup
key to enter the
TRIGGER
function menu.
2. Press
Type
and select
Serial
.
3. Press
Protocol
and select
SPI
.
4. Press
Trigger Setting
softkey.
Figure 40 SPI TRIG SET Menu
5.
Press the
Trigger Type
softkey to select the trigger condition.
Table 2 Menu Explanations of the SPI trigger type
Function
Settings
Explanation
Menu
MISO
Master-In, Slave-Out
Trigger Type
MOSI
Master-Out, Slave-In
6. Press the
Data Length
softkey, and turn the Universal Knob to set the length of a
data. The range of data length is 4 to 96 bits.
7. Set the value of the trigger data.
Set the value of a bit:
a. Press the
Bit Roll
softkey to select a bit in data.
b. Press the
Bit Value
softkey to set the value of the selected bit.
Set the value of all bits:
a. Press the
All Same
softkey to set the value of all bits.
Table 3 Menu Explanations of the SPI Bit value
Function
Settings
Explanation
Menu
0
High voltage level
Bit Value
1
Low voltage level
X
Don’t care the voltage level
8. Press the
Next Page
softkey.
9. Press the
Bit Order
softkey to set the bit order(MSB or LSB).
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...