Scientific
SMO1000E
124/185
To Save REF Waveform to Internal Memory
Do the following steps to save the REF waveform to internal memory:
1. Press the
REF
button on the front to enter the REF WAVE function menu. Note that
when the time horizontal format is in X-Y mode, REF function cannot be enabled.
2. Press the
Source
softkey; then, turn the
Universal Knob
to select the source of
reference channel. The source includes analog channel and math waveforms.
3. Press the
Location
softkey; then, turn the
Universal Knob
to select the position to
save the REF waveform. The source includes analog channel and math waveforms.
4. Press the
Save
softkey to save the channel or math waveform to the appointed
location. The vertical scale information and the vertical offset of the waveform will be
saved at the same time. It will pop out the message “
Store Data Success
” when
the waveform has been saved successfully.
Note: The REF waveforms are non-volatile. The REF waveform can still be
saved after restarts or default operation.
To Display REF Waveform
To the following steps to display REF waveform:
1. Press the
REF
button on the front to enter the REF WAVE function menu.
2. Press the
Location
softkey; then, turn the
Universal Knob
to select the
REF waveform that you want to display.
3. Press the
Display
softkey to select
On
to display the REF waveform on the screen.
Only saved location can be displayed. The oscilloscope can display all four reference
waveforms at a time.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...