184/185
6.
The display of waveform is ladder-like:
1) The horizontal time base might be too low. Increase the horizontal time base
to increase the horizontal resolution and improve the display.
2) If the display
Type
is “Vectors”, the lines between the sample points may cause
ladder-like display. Set
Type
to “Dots” to solve the problem.
7.
Fail to connect PC through USB:
Check the
IO Setting
in
Utility
to make sure whether the setting in
USB Device
matches the device currently connected. If needed, restart the oscilloscope.
8.
The USB storage device cannot be recognized:
1) Check whether the USB storage device can work normally.
2) Make sure whether the USB interface can work normally.
3) Make sure that the USB storage device being used is flash storage type. This
oscilloscope does not support hardware storage type.
4) Restart the instrument and then insert the USB storage device to check it.
5) If the USB storage device still cannot be used normally, please contact
your sales.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...