7. Press the
Display
softkey to select
Split
or
Full Screen
display mode.
Split:
the source channel and the FFT operation results are displayed
separately.
The time domain and frequency domain signals are displayed clearly
Full Screen:
the source channel and the FFT operation results are displayed in
the same window to view the frequency spectrum more clearly and to perform
more precise measurement.
Figure 43 FFT Waveform In Split Mode
Note:
Signals with DC components or deviation would cause error or deviation of the
FFT waveform components. To reduce the DC components, set the Channel
Coupling
to
AC.
To reduce the random noise and aliasing frequency components of repetitive or
single pulse, set the
Acquisition
of the oscilloscope to
Average
.
To measure FFT waveform:
To make cursor measurements, press the
Cursors
button, and then press the
Mode
softkey to select
On
to turn the cursors, Use the X1 and X2 cursors to measure
frequency values and the difference between two frequency values (
∆
X). Use the Y1 and
Y2 cursors to measure amplitude in dB and difference in amplitude (
∆
Y).
You can find the frequency value at the first occurrence of the waveform maximum by
using the X at Max Y measurement.
Note: please refer to the cursors chapter to obtain the method of using cursors.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...