Scientific
SMO1000E
45/185
Use the Zoom Function
Zoom is a horizontally expanded version of the normal display. You can use Zoom to
locate and horizontally expand part of the normal window for a more detailed (higher-
resolution) analysis of signals.
Press the
HORIZONTAL Scale Knob
on the front panel to turn on the zoom function,
and press the button again to turn off the function. When Zoom function is on, the display
divides in half. The top half of the display shows the normal time base window and the
bottom half displays a faster Zoom time base window.
Normal time base
Zoom time base
The area of the normal display that is expanded is outlined with a box and the rest of the
normal display is ghosted. The box shows the portion of the normal sweep that is
expanded in the lower half.
To change the time base for the Zoom window, turn the
Horizontal Scale Knob
. The
Horizontal Scale Knob
controls the size of the box. The
Horizontal Position Knob
sets
the left- to- right position of the zoom window. The delay value, which is the time
displayed relative to the trigger point is momentarily displayed in the upper- right corner
of the display when the
Horizontal Position Knob
is turned. Negative delay values
indicate you're looking at a portion of the waveform before the trigger event, and positive
values indicate you're looking at the waveform after the trigger event.
To change the time base of the normal window, turn off Zoom; then, turn the
Horizontal
Scale Knob
.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...