To Set Color Display
Color temperature adopts the change of waveforms’ color to reflect the change of the
waveforms’ appearing probability. The greater the probability that the waveform appears, the
warmer the color is; the smaller the waveform appears, the colder the color is.
The picture below shows the change of color from cold to warm. Press the Display button
on the front panel, and then press the
Color
softkey and set the option to
On
to turn on
the color temperature function. You can compare the waveform’s color with the picture
below to just the probability that the waveform appears.
Cold Color
Warm Color
Figure 58 Color Temperature
To Set and Clear Persistence
With persistence, the oscilloscope updates the display with new acquisitions, but does
not immediately erase the results of previous acquisitions. All previous acquisitions are
displayed with reduced intensity. New acquisitions are shown in their normal color with
normal intensity.
Do the following steps to set and clear persistence:
1. Press the
Display/Persist
button on the front panel to enter the DISPLAY function
menu.
2. Press the
Persist
softkey; then turn the
Universal Knob
to select the desires option.
Off
—turn of persistence.
Variable persistence time (1 second, 5 seconds, 10 seconds, 30 seconds ) — select
different persistence time, the results of previous acquisitions are erased after a
certain amount of time
Infinite
—select “Infinite” Results of previous acquisitions are never erased. Use
infinite persistence to measure noise and jitter, to see the worst- case extremes of
varying waveforms, to look for timing violations, or to capture events that occur
infrequently.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...