To Make Cursor Measurements
1. Press the
Cursors
button on the front panel to enter the CURSOR function menu.
2. Press the
Mode
softkey and set the option to
On
.
3. Press the
Source
softkey, and then use the
Universal Knob
to select the desired
source. Only analog channels ( CH1,CH2), math waveforms and reference
waveforms (REFA, REFB) that are displayed are available for cursors.
4. To make cursor measurements:
To measure the horizontal time, use the
Universal Knob
to move the X1 and X2
cursors to desired place. If necessary, set the cursor type to
X2-X1
, move X1
and X2 cursors together.
To measure vertical voltage or current, use the
Universal Knob
to move the Y1
and Y2 cursors to desired place. If necessary, set the cursor type to “Y2-Y1”,
move Y1 and Y2 cursors together.
To adjust the transparence of the cursors message box, press the
Display/Persist
button and go to the second page, press the
Transparence
(20% to 80%) softkey and then turn the
Universal Knob
to adjust the
transparence to the desired value.
Cursor examples:
1. Use cursors to measure pulse width:
Figure 48 Measure Pulse Width
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...