![Scientific SMO1000E Скачать руководство пользователя страница 165](http://html.mh-extra.com/html/scientific/smo1000e/smo1000e_user-manual_1217342165.webp)
Scientific
SMO1000E
165/185
To Do Pass/Fail Test
One way to verify a waveform's compliance to a particular set of parameters is to use
pass/fail testing. A pass/fail defines a region of the oscilloscope's display in which the
waveform must remain in order to comply with chosen parameters. Compliance to
the mask is verified point- by- point across the display.
The test results can be displayed on the screen as well as be declared through the
system sound or the pulse signal output from the
[TRIG OUT]
connector at the rear
panel. Pass/Fail test operates on displayed analog channels; it does not operate on
channels that are not displayed.
Figure 66 Pass/Fail Test
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...