Scientific
SMO1000E
119/185
LIN Triggering
This part will provide a brief introduction and description for the operation of the LIN
trigger.
To introduce the trigger conditions
Break
— The oscilloscope will be triggered at the position of break field’s break
delimiter.
ID
(Frame ID) — The oscilloscope will be triggered at the position of identifier field’s
stop bit, if the value of a frame’s ID is equal to specified value.
(Note: If the data’s value is 0xXX, any data value will be matched)
ID + Data
(Frame ID and Data) — The oscilloscope triggers when a frame with an ID
and data equal to the selected values is detected. Use the Universal Knob to select
the value for the ID, Data1 and Data2.
1) The ID’s value is the same as set value.
2) If you have set either Data1’s or Data2’s value, and the signal has a data is
the same as that value. If you have set both Data1’s and Data2’s value, the
signal should has two consecutive data, the first data’s value is Data1,
second data value is Data2.
(
Note
: If the data’s value is 0xXX, any data value will be matched)
Data Error
—The oscilloscope will be triggered when errors(such as ID check error,
checksum error, sync byte field error) are detected.
Operation steps:
1. Press
Setup
to enter the
TRIGGER
function menu.
2. Press
Type
and select
Serial
.
3. Press
Protocol
and select
I2C
.
4. Press
Trigger Setting
softkey to enter
LIN TRIG SET
menu.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...