Scientific
SMO1000E
100/185
SPI Triggering and Serial Decode
Please in order of “
Setup for SPI Signals
”, “
SPI Triggering
” and “
SPI Decode
” to
trigger and decode the signals.
Setup for SPI Signals
Setting the SPI (Serial Peripheral Interface) signal includes two steps: connecting the
CLK, MISO, MOSI and CS signals to oscilloscope; specifying the parameters of each
input signal.
1. Press the
Decode
key to enter the
DECODE
function menu.
2. Press the
Decode
softkey and select the desired slot (Decode1 or Decode2).
3. Press
Protocol
softkey and then select
SPI
by turning Universal Knob.
4. Press
Signal
softkey to enter the
SIGNAL
menu as Figure 35 shows.
Figure 36 SINGAL Menu
5. Set CLK (clock signal):
a. Press the
CLK
softkey to enter
CLK
menu.
b. Press the
CLK
softkey to select the channel that is connected to the SPI clock
signal.
c. Press the
Threshold
softkey to set the SPI clock signal’s threshold voltage level
by Universal Knob. The threshold voltage level is for decoding, and it will be
regard as the trigger voltage level when set the trigger type to serial.
d. Press the
Edge Select
softkey to set the oscilloscope will samples at clock
signal’s rising edge or falling edge.
e. Press
softkey to return previous menu.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...