Runt Trigger
The Runt trigger looks for pulses that cross one threshold but not another as shown in
the picture below.
Positive runt pulse
High level
Low level
Negative runt pulse
A positive runt pulse across through a lower threshold but not an upper threshold.
A negative runt pulse across through an upper threshold but not a lower threshold.
To trigger on runt pulse:
1. Press the
Setup
button on the front panel to enter the
TRIGGER
system function
menu.
2. Press the
Type
softkey; then turn the
Universal Knob
to select
DropOut
and
push down the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select CH1 or CH2 as
the trigger source.
4. Press the
Polarity
softkey to select
Positive
or
Negative
pulse to trigger.