(within a range of time value):trigger when the positive or negative pulse
time of the input signal is greater than the specified lower limit of time and
lower than the specified upper limit of time value.
For example, for a positive pulse, if you set t (pulse real width) >100ns
and t<300ns, the waveform will trigger.
--][--
(outside a range of time value): trigger when the positive or negative pulse
time of the input signal is greater than the specified upper limit of time and
lower than the specified lower limit of time value.
Figure 25 Pulse Trigger