Scientific
SMO1000E
59/185
Change the Horizontal Format
Press the
Acquire
button on the front panel; then press the
XY
soft key to set the XY(On)
or YT(Off) mode. The default setup is
YT
.
YT
It is the normal viewing mode for the oscilloscope. In the Normal time mode, signal
events occurring before the trigger are plotted to the left of the trigger point and signal
events after the trigger plotted to the right of the trigger point.
XY
XY mode changes the display from a volt- versus- time display to a volt- versus- volt
display. Channel 1 amplitude is plotted on the X- axis and Channel 2 amplitude is plotted
on the Y- axis, the two channels will be turned on or off together.
You can use XY mode to compare frequency and phase relationships between two
signals. XY mode can also be used with transducers to display strain versus
displacement, flow versus pressure, volts versus current, or voltage versus frequency.
The phase deviation between two signals with the same frequency can be easily
measured via Lissajous method. The figure below shows the measurement
schematic diagram of the phase deviation
According to
sin
θ
=A/B
or
C/D
(wherein,
θ
is the phase deviation angle between the two
channels and the definitions of A, B, C and D are as shown in the figure above), the
phase deviation angle is obtained, that is:
θ
=
±
arcsin (A/B) or
±
arcsin (C/D)
If the principal axis of the ellipse is within quadrant I and III, the phase deviation angle
obtained should be within quadrant I and IV, namely within (0 to
π
/2) or (3
π
/2 to 2
π
). If
the principal axis of the ellipse is within quadrant II and IV, the phase deviation angle
obtained should be within quadrant II and III, namely within (
π
/2 to
π
) or (
π
to 3
π
/2).
X-Y function can be used to measure the phase deviation occurred when the signal under
test passes through a circuit network. Connect the oscilloscope to the circuit to monitor
the input and output signals of the circuit.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...