Use Sequence Mode
Sequence is also a kind of acquisition mode, which does not display waveform during
sampling process. It improves the waveform capture rate, and the maximal capture rate
is more than 400,000 wfs/s. So it can capture the small probability event effectively.
The oscilloscope runs and fills a memory segment for each trigger event. The
oscilloscope is busy acquiring multiple segments. The oscilloscope continues to trigger
until memory is filled, and then display the waveforms on the screen.
To use the sequence mode, the HORIZONTAL Format must be set to
YT
.
Do the following steps to use the sequence mode.
1. Press the
Acquire
button on the front panel to enter the ACQUIRE function menu;
2. Press the
Sequence
softkey to enter the SEQUENCE function menu.
Figure 19 SEQUENCE Function Menu
3. Press the
Segments Set
softkey; and then turn the
Universal Knob
to select the
desired value.
Do the following steps to replay the sequence waveform under history mode:
1. Press the
History
softkey to enable HISTORY function .
Figure 20
HISTORY Function Menu
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...