Scientific
SMO1000E
83/185
Interval Trigger
Trigger when the times difference between the neighboring rising or falling edges
meets the time limit (<=, >=, [--,--], --][--).
Trigger
Trigger
To set interval trigger:
1. Press the
Setup
button on the front panel to enter the TRIGGER system function
menu.
2. Press the
Type
softkey; then use the
Universal Knob
to select
Interval
and push
down the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select CH1 or CH2 as
the trigger source.
4. Press the
Slope
softkey to select rising or falling edge.
5. Press the
Limit Range
softkey; turn the
Universal Knob
to select desired condition.
<=
(less than a time value): trigger when the positive or negative pulse time of
the
input signal is lower than the specified time value.
>=
(greater than a time value): trigger when the positive or negative pulse time
of
the input signal is greater than the specified time value.
[--,--]
(within a range of time value):trigger when the positive or negative pulse
time of the input signal is greater than the specified lower limit of time and
lower than the specified upper limit of time value.
--][--
(outside a range of time value): trigger when the positive or negative pulse
time of the input signal is greater than the specified upper limit of time and
lower than the specified lower limit of time value.
6. Press the
Time Setting
softkey (
<=, >=, [--,--],--][--
), turn the
Universal Knob
to select the desired value.
Содержание SMO1000E
Страница 1: ...Digital oscilloscope SMO1000E User Manual ...
Страница 17: ...Scientific SMO1000E 17 185 Appearance and Dimensions Figure 1 Front View Figure 2 Top View ...
Страница 54: ...Scientific SMO1000E 54 185 Figure 12 x Interpolation Figure 13 Sinx x Interpolation ...
Страница 82: ...Scientific SMO1000E 82 185 Figure 28 Relative Window Trigger ...
Страница 84: ...Scientific SMO1000E 84 185 Figure 29 Interval Trigger ...
Страница 90: ...Scientific SMO1000E 90 185 Universal Knob to select the desired value Figure 33 Pattern Trigger ...
Страница 109: ...Scientific SMO1000E 109 185 Figure 53 UART Trigger ...