Technical Data
12
User guide
Defibrillation Waveform
12.2
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
Standard energy settings
AED
Adult
Paediatric
Manual mode
Adult
Paediatric
Deviation at 50
3 J or
15 % (the higher value is assumed)
150/200/200 joules (configurable see page
)
50/50/50 joules (configurable see page
(automatic selected when the paediatric or adult electrodes are connected)
150
joules (configurable see page
50
joules (configurable see page
)
(default energy settings when starting in manual mode, adjustable anytime during
intervention)
Charging time for shock
• with fully charged battery
• with 15 VDC mains voltage after 15
discharges with max. energy
• from switch-on of the device with
pads
(Time used to charge the storage capacitor to the max. energy of 200 J in manual
mode)
8 seconds
9 seconds
19 seconds
Cycle Time Rhythm Analysis –
Shock Standby in AED Mode
• with fully charged battery
• with 15 VDC mains voltage after 15
discharges with max. energy
• from switch-on of the device to
charge at max. energy
1st shock = 11 s max.
1st shock = 11 s max.
1st shock = 23 s
Cycle time shock – shock
<15 s
Operating Modes
• Synchronised with heart action < 60 ms after R wave
• Non synchronised
• AED
Charge control and monitoring
• Automatic shock recommendation of analysis in AED mode
• Direct via touch screen
• Display of selected energy
Patient resistance
25 ...250
Indication when ready to shock
LED below
is lit
Shock delivery
Using key