10.4 Maintenance interval of the batteries
Page 118
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
10.4 Maintenance interval of the batteries
10.4.1
Replacing the batteries
Replacing the power battery:
The power battery needs to be replaced if the operating time in monitoring mode is
less than 1 hour with a fully charged battery (see chapter
Replacing safety primary cell:
Check in the menu if safety cell is OK. (see chapter
10.4.2
Battery disposal
Important
• The battery's performance and life largely depend on how and under which
ambient conditions the battery is used.
Power Battery
• The rechargeable power battery is maintenance-free during its normal life.
• The battery must be replaced according the expired date on the battery, regardless
of whether or not the unit has been used.
• Only store fully charged batteries. If a battery is not used, recharge it every 6
months.
Safety primary cell
• The safety primary cell is maintenance-free during its normal life.
• The safety primary cell must be replaced according the expired date on the Cell,
regardless of whether or not the unit has been used.
Danger of explosion! Battery may not be burned or disposed of with domestic
refuse.
Danger of acid burns! Do not open or heat up the battery.
The battery is to be disposed of in municipally approved areas or sent back to
SCHILLER Médical.