DEFIGARD/PHYSIOGARD Touch 7
4.10 CO2 mainstream
4.10.1
IRMA mainstream gas analyser
IRMA mainstream gas analyser is intended to be connected to a patient breathing
circuit for monitoring of inspired/expired gases during anesthesia, recovery and
respiratory care.
The IRMA gas analyzer is intended for use by authorized healthcare
professionals only.
Risk of cross infection!
– Disposable airway adapters must not be re-used.
– Used disposable airway adapters must be disposed of in accordance with local
regulations for contaminated and biologically hazardous fluids.
Airway Adapters are non-sterile devices. Do not sterilise.
Use only Masimo manufactured IRMA airway adapters.
Use the correct adapter:
– Do not use the IRMA Paediatric/Adult airway adapter for infants because the
adult adapter adds 6 ml dead space to the patient circuit.
– Do not use the IRMA Infant airway adapter for adults because this may cause
excessive flow impedance.
– Do not use the airway adapter with metered dose inhalers or nebulized medica-
tions as this may affect the light transmission of the airway adapter windows.
The IRMA sensor is not intended to be used as the only means of monitoring a
patient.
Never sterilise or immerse the sensor in liquid.
Do not apply tension to the sensor cable.
Do not place the IRMA airway adapter between the endotracheal tube and an
elbow as this may allow patient secretions to block the adapter windows and
result in incorrect operation.
To keep secretions and moisture from pooling on the windows, always position
the IRMA probe in a vertical position with the LED pointing upwards.
Replace the airway adapter if rainout/condensation occurs inside the airway
adapter.