Monitoring
4
User guide
Operator-Defined Alarm Thresholds
4.3
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
4.3.1
Table of wide/narrow threshold setting
The range values in brackets
()
are the default values activated when pressing “De-
fault” in the alarm setting menu see
The values before the bracket
“ECG 0-350
(50-150) bpm”
are the low/high system limits.
HR [bpm]
Pat. value
Wide limits
Narrow Limits
Range:
Low
High
Low
High
ECG 0-350
(50-150) bpm
<60
-20
+35
-10
+25
Pleth 25-240
(50-150) bpm
60-79
-25
+40
-20
+30
80-104
-30
+40
-30
+30
>=105
-35
+45
-25
+25
Allowed values
[30-150]
[100-240]
[30-150]
[100-240]
RR [rpm]
Pat. value
Wide limits
Narrow Limits
Range:
0-60
(5-30) resp/min
Low
High
Low
High
<15
-8
+8
-4
+4
>=15
-15
+15
-8
+8
Allowed values
[5-15]
[10-60]
[5-15]
[10-60]
SpO2 [%]
Pat. value
Wide limits
Narrow Limits
Range: 50-100
(85-100) %
Low
High
Low
High
>=90
-5
+3
-5
+3
<90
-5
+3
-5
+3
Allowed values
[85-100]
[90-100]
[85-100]
[90-100]
SpCO/SpMet [%]
Pat. value
Wide limits
Narrow Limits
Range:
Low
High
Low
High
SpCO 0-40
(0-10) %
-
0%
10%
0%
10%
SpMet
0-15%
-
0%
3%
0%
3%
Temperature [°C]
Pat. value
Wide limits
Narrow Limits
Range: 0-46
(35-39) °C
Low
High
Low
High
-
-3
+3
-1
+1
Allowed value
-
[31-41]
[31-41]
[31-41]
[31-41]
EtCo2 [mmHg]/ [%]
Pat. value
Wide limits
Narrow Limits
Range: 5-70 mmHg
/
0.7-9.2 %
(15-50 mmHg/2-6.6 %)
Low
High
Low
High
<40/5.3
-10/1.3
+15/+2.0
-10/-1.3
+15/+2.0
>=40/5.3
-10/1.3
+15/+2.0
+15/+2.0
+15/+2.0
Allowed values [mmHg]/ [%]
[5-60] /
[0.7-7.9]
[20-70] /
[2.7-9.2]
[5-60] /
[0.7-7.9]
[20-70] /
[2.7-9.2]