![Schiller DEFIGARD Touch 7 Скачать руководство пользователя страница 98](http://html1.mh-extra.com/html/schiller/defigard-touch-7/defigard-touch-7_user-manual_1212081098.webp)
5
Defibrillation
5.7
Defibrillator Technical Messages
Page 98
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
5.7 Defibrillator Technical Messages
Alarm
Code
Cause
Remedy
Defibrillator inoperative
T.ECG11
T.DEFI01
T.DEFI02
T.DEFI03
T.DEFI04
T.DEFI05
T.DEFI06
T.DEFI07
T.DEFI10
T.DEFI13
T.DEFI14
T.DEFI15
• CPU peripheral board defective
• Defi board defective
Contact technical service
Incorrect electrodes
T.DEFI08
T.DEFI12
• Internal discharge while releasing
shock “on impedance higher than
250 ohm” or if the current during the
shock is 0 or above 105 A, because
of poorly placed electrodes
Check electrodes, if necessary re-
apply electrodes
Internal discharge duration
too long
T.DEFI09
• A shock that is not delivered within
the specified duration causes an in-
ternal discharge.
Do not exceed the time of 20 sec-
onds till releasing the shock
Contact technical service
Internal discharge smaller
energy
T.DEFI11
• Internal discharge because of select-
ing lower energy after the device
charged the selected higher energy
Normal safety discharge