![Schiller DEFIGARD Touch 7 Скачать руководство пользователя страница 154](http://html1.mh-extra.com/html/schiller/defigard-touch-7/defigard-touch-7_user-manual_1212081154.webp)
12
Technical Data
12.6 Device Configuration
Page 154
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
12.6.18
WIFI
12.6.19
GSM
12.6.20
SEMA
Parameter
Values
Description
SSID
-
SSID of the Wi-Fi Network that shall be used for data transmission
Encryption type
• WEP
• WPA
• WPA2*
• WPA-EAP
• WPA2-EAP
• None
Type of encryption from the Wi- Fi Network that shall be used for data
transmission
Security key
-
Security key of the Wi-Fi Network that shall be used for data
transmission
Wi-Fi Login
-
Login used for EAP authentication
Wi-Fi Password
-
Password used for EAP authentication
Wi-Fi ping server
-
Sets the IP that will be used to test the connectivity of the device over
WIFI. Usually, it's recommended to use either common DNS
server(e.g.
8.8.8.8), or the public SEMA Server IP address
Parameter
Values
Description
PIN
-
Sets the PIN code to unlock SIM card
It's recommended to either use the same PIN for all devices, or to
disable PIN on all SIM cards. Both are accepted by the device
APN name
-
Sets the APN of the M2M Internet provider
Example: orange.m2m.spec or MATOOMA, or A1.net
APN port
-
Sets the APN port (if required by the APN)
APN user
-
Sets the APN user (if required by the APN)
APN password
-
Sets the APN password (if required by the APN)
APN extra
commands
-
If required by the APN, extra commands can be specified here
GSM ping server
8.8.8.8
Sets the IP that will be used to test the connectivity of the device over
GSM. Usually, it's recommended to use either common DNS
server(e.g. 8.8.8.8), or the public SEMA Server IP address
Parameter
Values
Description
SEMA server
-
Sets the IP address or the hostname of the SEMA Server
Should the device be able to send data to SEMA over Internet (typically
through GSM), the public IP address or hostname must be specified
and not the private
Example: 188.165.287.137
(public IP) and not 192.168.200.32 (private IP)
SEMA server port
8181
Sets the port with which SEMA Server is reachable. This port might be
different whether the device tries to connect with or without SSL
encryption
Example: By default, 8080 is used for HTTP and 8181 is used for
HTTPS
SEMA SSL
encryption
• False
• True*
Sets SSL encryption. It's recommended to activate this option to
secure the communication, especially if transmission are done over the
Internet