Page 127
SCHILLER Charging Unit CS-1
11
User guide
Battery Charging Options
11.1
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
11 SCHILLER Charging Unit
CS-1
11.1 Battery Charging Options
The following options are available to charge the batteries:
• Batteries can be removed and charged using the optional charging unit SCHILLER
CS-1 (see following).
• The DEFIGARD/PHYSIOGARD Touch 7 batteries are also charged when the
DEFIGARD/PHYSIOGARD Touch 7 is connected to the external power supply.
11.2 Inserting a battery
• Insert the battery in the charger unit and push home until the battery clicks in place
with the blue catches.
• To remove a battery, press the two blue catches to release it.
(1) Control/indicator panel
(2) Battery slot 1 - only this slot can be used for battery calibration.
(3) Battery slot 2.
The batteries supplied are rechargeable Lithium-Ion 4.65 Ah batteries. Only use
rechargeable batteries supplied by SCHILLER.
We recommend that the batteries are replaced every 1000 charge / discharge
cycles.
In battery slot 1, the battery can also be calibrated.
The charging unit is optional.
3
2
1
The charger has double contacts and a battery can be inserted in either way.