SCHILLER Charging Unit CS-1
11
User guide
Battery Calibration
11.4
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
11.4 Battery Calibration
Every battery has an individual calibration and capacity circuit. The DEFIGARD/
PHYSIOGARD Touch 7 uses this information to display the battery capacity. New
batteries are factory calibrated before use and should not need recalibrating during
their normal life cycle. If a battery seems to have a low capacity or is near the end of
its life, it may need recalibrating.
Calibrate a battery as follows:
1. Place the battery in the left battery slot of the charger.
2. Press the
Cal
key to start the calibration procedure. The stages to calibrate a bat-
tery are as follows:
– The battery is completely discharged.
1
– The battery is charged and calibrated.
The calibration cycle takes approximately 2.5 - 5 hours to complete - the full charge
cycle (after full discharge) takes approximately 2.5 hours. During the calibration
process:
– The LED above the calibration key is lit.
– To indicate the discharge cycle the three indicators LEDs blink in sequence -
right to left.
– When the discharge cycle is finished, the three indicators LEDs blink in se-
quence left to right to indicate the charge cycle.
– When the calibration is successfully completed, all LEDs are lit and the LED
above the calibration key is extinguished.
If for any reason the battery cannot be calibrated, the LED above the
CAL
key is
flashing. This indicates that the battery cannot be calibrated and the battery should
not be used any longer.
1. The power stored in the battery to be calibrated is transferred by ‘dynamic pow-
er distribution’ to the battery in the other slot, or used for any connected device,
so that no energy is wasted.
The calibration process can be stopped at any time, by again pressing the
Cal
key.