Pacemaker
6
User guide
Start-up of the Pacemaker
6.4
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
6.4.1
Pacemaker display
6.4.2
Selecting pacemaker mode
1. Press the Pacer measurement field
(1) to open the Pacer menu (2).
2. Press operational mode
Fix
or
Demand (3)
3. The operational mode is displayed in the pacer measurement field
(1)
4. When the pacemaker is running, press
Close (4)
or switch to monitoring with the
monitoring key.
In this case, the pacemaker values are displayed as small measurement field
(1)
at
the top right and the curve field shows again all curves.
• Select the
Pacer
measurement field at the top right to display the pacemaker
function.
The pacemaker menu with the pacemaker parameter is displayed.
The pacemaker default mode at switchover is “
Demand
” mode; the “
Fix
” mode has
to be selected manually.
1
2
3
4
Type of used
electrodes
Running time
of pacer
Impedance OK/
High/Low
Pacer pulse
Frequency
Current
Fixed
Demand
HR
Screenshot
Close
Manual def
Event
Start