![Schiller DEFIGARD Touch 7 Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/schiller/defigard-touch-7/defigard-touch-7_user-manual_1212081088.webp)
5
Defibrillation
5.3
Manual Defibrillation Using Pads
Page 88
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
5.3.4
Manual Defibrillation Using Pads Procedure
1. Connect the electrode cable to the pads connector.
2. If the device starts in
Monitoring
or
AED
mode, proceed according to the descrip-
tion in
chapter 5.2.1 Activating the manual defibrillation mode, page 82.
3. Select the energy via the touch -.
4. Initiate the energy charging by pressing
"Charge"
.
Danger of electric shock!
• Do not, under any circumstances, touch the patient during shock delivery.
• Make sure that the patient does not touch any conducting objects.
5. Trigger the shock by pressing the button
.
6. Finish the therapy (see
7 Finishing the Therapy page 106
Fig. 5.5
Manual defibrillation
Event
CPR Guide
Metronome