DEFIGARD/PHYSIOGARD Touch 7
4.2 Alarm System
4.2.1
Alarm priority
4.2.2
Operator’s position
The visual alarm LED is visible to a distance of 4 meters and the flashing value is
visible to a distance of 1 meter.
4.2.3
Alarm list
An alarm list can be displayed any time by touching the alarm status line.
In some countries, it is not permitted to disable audio alarms
permanently. Therefore, this function is configurable.
When pausing or switching off the audio alarm, even high-priority alarms such
as VT/ VF and asystole are paused/switched off!
Pausing or switching off of the audio alarm system is only allowed if the patient
is permanently observed.
Alarm type
Priority
Audible signal
Display
Technical alarm
Low
One beep once
• Text display in the alarm status field at the top
• Displaying -?- in the parameter field
• Orange LED is lit
Physiological alarm
Medium
3 signals (beep beep beep)
• Text display in the alarm status field at the top
• Orange flashing parameter field
• Orange LED is flashing
Physiological alarm
High
10 signals
(beep beep beep - beep
beep beep beep beep - beep
beep)
• Text display in the alarm status field at the top
• Red flashing parameter field
• RED LED is flashing
Ensure that the environmental noise is below the alarm sound volume of 65 dB.
Alarm List
08h02m43s: HR out of range (P-ECG03)
08h02m43s:SpO2: Sensor Off Patient (T-SP229)
HR out of range