Technical Data
12
User guide
System data
12.1
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
Memory
24 hours memory (FIFO)
Recording of Defi, ECG Lead II, Impedance curves, Events, CPR feedback,
patient data, patient vitals, screenshots
Safety standard
IEC/EN 60601-2-4
The device is designed for intensive use
EMC
• IEC/EN 60601-1-2
• IEC/EN 60601-2-4
• CISPR 11 class B
The device can be exposed to the following interferences without any impairment:
• Static discharges up to 15 kV
• Field strength up to 20 V/m in the radio frequency range of (80...2500 MHz, 5 Hz
modulated)
• Magnetic fields of 100 A/m, 50 Hz
Conformity
CE according to directive 93/42/EEC class IIb
Protection class
Class I according to IEC/EN 60601-1