4
Monitoring
4.6
SpO
2
-
,
SpCO
,
SpMet monitoring (Option)
Page 54
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
Check sensor
T.SP216
T.SP217
T.SP218
T.SP219
T.SP220
T.SP221
T.SP222
T.SP223
T.SP224
T.SP225
T.SP227
• No sensor connected
• Sensor life expired
• Incompatible sensor
• Unrecognised sensor
• Cable and sensor fault
• No adhesive sensor connected
• Adhesive sensor life expired
• Incompatible adhesive sensor
• Unrecognised adhesive sensor
• Defective adhesive sensor
• Defective sensor
Check/replace the sensor
Check cable and sensor fault
Alarm (measurement field) Code
Cause
Remedy