4
Monitoring
4.10 CO2 mainstream
Page 64
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
4.10.2
Preparing the IRMA sensor
1. Insert the airway adapter.
2. Connect the 15 mm airway adapter end to the ventilator Y piece.
3. Connect the patient side of the airway adapter to the tube.
IMPORTANT
4. To keep moisture from pooling on the windows, position the airway adapter with
its windows in a vertical position.(LED pointing upwards)
To prevent the windows' soiling by secretions of the patient or condensing water,
position the adapter in a slightly angled position between the endotracheal and
the respiration tubes.
Alternatively; connect a HME (Heat Moisture Exchanger) between the patient’s
endotracheal tube and the IRMA probe. Placing a HME in front of the IRMA probe
protects the airway adapter from secretions and effects of water vapor and
eliminates the need of changing the adapter. It allows free positioning of the IRMA
probe as well. Unless the IRMA probe is protected with a HME always position
the IRMA probe with the LED pointing upwards.
1
2
3
4
adapter window
HME