Technical Data
12
User guide
Device Configuration
12.6
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
12.6.8
NIBP
12.6.9
SpO2
12.6.10
Temp
12.6.11
EtCO2
Parameter
Values
Description
Deflation rate
• 3, 4, 5, 6,
7
, 8, 9 mmHg/s
Sets the cuff deflation rate
NIBP unit
• mmHg*
• kpa
Sets the unit in which the NIBP values will be displayed and stored
Automatic cycles at
start
•
False
*
• True
Automatic NIBP measurement cycle starts once a first measurement is
initiated manually
Initial pressure for
adults
• 90, 120, 150,
180*,
210,
240, 270mmHg
Sets the initial cuff pressure for measurement in adult mode
Initial pressure for
children
• 90, 120,
150*
, 180, 210,
240, 270mmHg
Sets the initial cuff pressure for measurement in child mode
Initial pressure for
neonate
•
50*
, 70, 90 110,130,
150mmHg
Sets the initial cuff pressure for measurement in neonate mode
Parameter
Values
Description
SpO2 average
• 4, 6, 8, 10,
12*,
14, 16sec
Sets the integration time for the calculation of the displayed average
value.
SpO2 sensitivity
• Normal*
• Automatic Probe Off De-
tection
Sets the measurement sensitivity.
Adaptive Probe Off Detection is optimised for the detection of Sensor
has come off, regardless of the signal quality.
SpO2 sound level
• Off
• Low
• Medium*
• High
Sets the audio level of the pulse sound
Parameter
Values
Description
Temperature unit
• Celcius*
• Fahrenheit
Sets the unit in which the temperature value will be displayed and
stored
Parameter
Values
Description
Respiration curve
amplitude
•
50*,
75, 100mmHg
Sets the respiration curve amplitude to be displayed by default
EtCO2 unit
• vol%
• mmHg*
Sets the unit in which the EtCO2 value will be displayed and stored