11
SCHILLER Charging Unit CS-1
11.3 Control Panel
Page 128
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
11.3 Control Panel
The LED indicators give the following information:
Power connected
indicator
Battery slot 1
indicators
Calibration button
Battery slot 2
indicators
LED
Function
No LEDs lit
- No battery inserted.
Constant display for 5 seconds
- When a battery is first
inserted, one, two or three LEDs are lit for 5 seconds.
This indicates the battery charge state (1 LED = 1/3
capacity, 2 LEDs = 2/3 capacity and 3 LEDs = 100%
capacity).
LEDs are flashing in sequence (left to right)-
After 5
seconds of a battery being inserted, the charge
sequence starts.
All LEDs lit
- battery fully charged.
LEDs are flashing in sequence (right to left) -
battery
LEDs blink two/one in sequence
- faulty battery.